DESCRIPTION
Infant Resuscitator
- It is mainly used on delivery room, baby wards, NICU that the baby’s maximum weight up to 10KG.
- Inside an alarm system of differential pressure. When alarming, the safety valve starts automatically and produces the mixture gas continuously to ensure the safety of patients.
- Adjust the Oxygen concentration from 21% to 100% optionally
- The range of flow adjustment is 5LPM~15LPM.
- The MAX-P and PIP are safe, controlled and accurate.
- The PEEP is constant and accurate.
- Simple operation and flexible application.
- The output pressure would not be influenced by the operator experience, training, attention and fatigue.
Technical parameter:
- Pressure gauge : -2~10kPa (-20~100cmH2O).
- MAX-P : @0~15LPM ≤6kPa (60cmH2O).
- PIP : @5~15LPM 0.2~5.7kPa (2~57cmH2O).
- PEEP : @5~15LPM 0.03~2.3kPa (0.3~23cmH2O).